Glove Box Regeneration Measures

De WIKI ISFODOSU
Ir a la navegación Ir a la búsqueda

1) Regeneration gasoline: a mixture of hydrogen and inert gasoline (hydrogen accounts for 5% -10%)

two) Quantity: 40L gasoline cylinder 10Mpa

3) Regenerative gasoline stress: .05Mpa (the gas consumption for each 50 percent hour in the state of .05Mpa is that the cylinder pressure drops by one.5Mpa. If it exceeds one.5Mpa, the force of the force regulator’s schedule will be diminished, if it is much less than one.5Mpa Increase the tension of the strain regulator routine)

4) Process: three hrs heating-3 hrs heating and vac glovebox air flow-three hrs vacuuming-one hour pumping-ten hours cooling

5) Methods:

① Hook up the regeneration fuel (check the pipeline for leaks soon after relationship: after connecting, open the principal valve of the force reducing valve to the optimum, open the auxiliary meter to .1Mpa, switch off the main valve, hold out 15 minutes, if the tension of the auxiliary meter continues to be at .one Mpa stays unchanged, indicating no air leakage

>②Set the force of the auxiliary gauge of the force cutting down valve to .1Mpa (the tension through ventilation is about .06-.08MPa

>③ Close the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Begin regenerat

br>⑥ The new program ends at 20h (clear ahead of cleansing, water and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.